In this case there has been a problem  with the loading of the program from the PC to the HHR.

Turn the HHR off. Connect the HHR to the PC with the USB cable (do not connect the power cable). Make sure the HHR is turned off (black screen), if it is turned on hold the “on/off” button to turn it off. Load program to HHR again by clicking “start” on the Program Loader. Turn the HHR on while the Program Loader is “looking for device”. Make sure all kables are well connected. This should solve your problem.